Detection level monitor
This screen enables to display the detection level and the values on the detection area at a
real time basis. When entering to this screen, the contact output condition is reset.
Fig. 50 Detection level monitor screen
Forcible output of current
In this screen, current output is available with the specified value.
Current output condition is held as entering into this screen.
Fig. 51 Forcible output of current screen
Operation
Follow the step below for forcible output of current:
1.
Set the output level by touching
/
keys.
2.
Touch “OUT”; the value set at the step 1. is used for current output.
Name
Description
Detection level
Displays the detection level in the unit of 0.1 mA.
Detection area
Displays the occupied rate of oil film against detection area in the unit of 10
.
Contact output
Selecting “ON” enables contact output at oil film alarm.
Selecting “OFF” (default) does not enable contact output even at oil film alarm. This
setting is valid only in this screen.
Contact output
Detection level
Detection area
Name
Description
Output level
Output level: 3 mA to 21 mA; changeable in the unit of 0.1 mA
Output level