![horiba LO-300 Instruction Manual Download Page 63](http://html1.mh-extra.com/html/horiba/lo-300/lo-300_instruction-manual_2162291063.webp)
Operation
49
History
This function enables to display as many as 100 histories from the latest one.
Fig. 48 History screen
Name
Description
Page
Shows the current page. 10 histories are displayed per page.
History
Year, Date, Time, and status are displayed. The type of status is as follows.
ALRM: Oil film detection (alarm)
WARM: Oil film detection (warning)
HIGH: Upper limit error
LOW: Lower limit error
NORM: Normal return (from oil film alarm, oil film warning, upper limit error, lower limit
error)
TEMP: Temperature error
MTR: Motor stop, Frequency of 40 Hz or less, Luminescence stop
REV: Motor rotation error
LDL: Laser replacement notification
MEM: Flush ROM error; successful rewrite
MEME: Flush ROM error; failure in rewrite
BACK
Displays the previous page.
NEXT
Displays the following page.
ESC
Returns to the monitor display.
Page
Summary of Contents for LO-300
Page 1: ...Laser scanning type surface oil detector LO 300 Instruction Manual CODE GZ0000312982B...
Page 12: ......