Technical Note
77
Conductivity Standard Values at Various Temperatures
Temp. (°C)
Conductivity value at 25°C
84 (μS/cm)
1413 (μS/cm)
12.88 (S/cm)
111.8 (mS/cm)
15
68
1147
10.48
92.5
16
70
1173
10.72
94.4
17
71
1199
10.95
96.3
18
73
1225
11.19
98.2
19
74
1251
11.43
100.2
20
76
1278
11.67
102.1
21
78
1305
11.91
104.0
22
79
1332
12.15
105.9
23
81
1359
12.39
107.9
24
82
1386
12.64
109.8
25
84
1413
12.88
111.8
26
86
1440
13.13
113.8
27
87
1467
13.37
115.7
28
89
1494
13.62
117.7
29
90
1521
13.87
119.7
30
92
1548
14.12
121.8
31
94
1575
14.37
123.9