Conductivity Setup
44
● P1.6 Calibration Clear Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
key,
P1.3 A.CAL
appears.
5. Press the
key,
P1.4 T.CFF
appears.
6. Press the
key,
P1.5 T.REF
appears.
7. Press the
key,
P1.6 C.CLR
appears.
8. Press the
ENT
key, default
NO
screen appears.
9. Use the
keys to change to
YES
. This erases the calibration data.
10. Press the
ENT
key,
CCLR DONE
screen appears briefly.
11.
P1.6 C.CLR
screen appears. This indicates completion of calibration clear setup.
key
key
key
key
key
key