Error Messages and Troubleshooting
73
Swelling of keypad
Part of the display is missing
Cause
How to solve problem
Using the instrument at high elevation
or other location where the air pressure
is different from sea level.
To eliminate the pressure difference between the
inside and outside of the instrument, briefly open and
then close the serial connector cover. After opening,
correctly close the cover to maintain dust and water
proofing.
Instrument defect.
Consult your dealer.
Cause
How to solve problem
Instrument defect.
Check the display by switching ON the instrument
when all the LCD segments are lit.