pH Calibration
12
Manual Calibration
1. After placing the pH electrode in pH buffer, press the
CAL
key.
2. The selected
CUST
pH buffer group appears on the meter screen.
3. Wait for the
to stabilize. When it stabilizes, there will be a sound indicating that
reading is stable.
4. Press
and
keys to enter the pH buffer value.
5. Press the
ENT
key to confirm and save calibration data.
6. Meter displays
CAL DONE
indicating end of the pH calibration procedure. Electrode
icon and number of calibration points done (e.g. 1 PT for 1 point, 2 PTS for 2 points) light
up in measurement mode.
7. Repeat above steps for other calibration points as required.
To view and/or print calibration data, press the
key when you are in the
CAL
mode.The
display scrolls through the calibrated values, segment slope(s), average slope, and offset.
key
key
key
Note