pH Calibration
10
■ Calibration
This section describes the calibration procedures using LAQUA 2000 series benchtop meters
and electrodes.
● pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration using
PH2000, ION2000, or PC2000 meter, follow the procedure detailed below:
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the pH electrode.
• Prepare the pH buffers required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the pH buffer.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH 7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH 7 first.
• Default pH buffer group in buffer setup is
USA
. If you want to change to
NIST
,
NIST2
,
DIN
,
or
CUST
(Custom) refer to “P1.1 Buffer Setup” on page 29.
• For
USA
,
NIST
,
NIST2
, and
DIN
pH buffer groups, follow the Auto Calibration procedure.
For
CUST
, follow Manual Calibration procedure.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to “P1.4 Calibration Clear Setup” on page 32.
Note
Tip
Summary of Contents for LAQUA 2000 Series
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 54: ...Ion Setup 36 key key key key...
Page 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Page 98: ...Printout Summary 80 Salinity TDS...
Page 99: ...Printout Summary 81 Printer Format Data Log...
Page 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Page 101: ...Printout Summary 83 Conductivity Salinity...
Page 106: ...88...
Page 107: ......