TDS Calibration
21
● TDS Calibration
Total dissolved solids (TDS) is calculated from the measured conductivity value so TDS
calibration is not required. Once conductivity mode is calibrated, TDS values will be
recalculated accordingly.
Set the appropriate TDS curve. Available TDS curves in the meter are as follows:
-
LINR
(Linear KCl curve with adjustable factor from 0.40 to 1.00)
-
442
(Myron L 442 non-linear curve)
-
EN
(European environmental standard non-linear curve)
-
NACL
(non-linear salinity curve)
• The default TDS curve is
LINR
.
To set a desired TDS curve, refer to “P2.1 TDS Curve
Note