Ion Setup
35
● P1.2 Ion Electrode Type Setup
1. Press the
key,
P1 ION
screen appears.
2. Press the
ENT
key,
P1.1 UNIT
screen appears.
3. Press the
key,
P1.2 TYPE
screen
appears.
4. Press the
ENT
key, default
NH3
(Ammonia) ion electrode type appears.
5. Use the
keys to change the electrode type to
CA
(Calcium),
POT
(Potassium),
NO3
(Nitrate),
FL
(Fluoride),
CL
(Chloride), or
CUST
(Custom). If
CUST
is selected,
default
CHRG
1
screen appears after pressing
ENT
key. Use
keys to
change to
-1
,
2
, or
-2
depending on the valence of the ion to be measured.
6. Press the
ENT
key,
P1.2 TYPE
screen appears indicating completion of ion
electrode type setup.