Conductivity Setup
39
● P1.1 Cell Constant Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
ENT
key, default
CELL 1.0000
appears.
4. Use the
and
keys to set the cell constant in between 0.0700 to 13.000.
Use the
key to move the decimal point.
5. Press the
ENT
key,
P1.1 CELL
screen appears. This indicates completion of cell
constant setup.