Temperature Calibration
24
● Temperature Calibration
Temperature calibration is required to accurately match electrode to the meter. Check the
temperature reading and if it is acceptable, no temperature calibration is required. If you need
to calibrate, please follow the procedure detailed below:
Prerequisites
• Clean the electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the electrode and temperature sensor.
• Dip the electrode in a solution with known temperature until its temperature sensor is
immersed.
• Wait for 5 minutes to ensure temperature stability.
• Meter displays
MTC
if the temperature sensor is not plugged in and displays
ATC
if the
temperature sensor is plugged in.
• Temperature calibration must be performed using a solution with known temperature or
against a calibrated thermometer.
To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip