Conductivity Setup
38
● P1 COND Setup
Using conductivity setup, you can:
• Set cell constant
• Select conductivity unit
• Select calibration method
• Set temperature coefficient
• Set reference temperature
• Erase calibration data
To set the conductivity function of EC2000 or PC2000 meter, follow the procedure detailed
below:
Prerequisites
• Switch on the meter.
• Ensure that the meter is in conductivity measurement mode.
• Default cell constant value is
1.0000.
You can set a value in between
0.0700 to 13.000
.
• Default conductivity unit is
S/cm
. You can change the unit to
S/m
.
• Default auto calibration setup is
ON
. To perform manual calibration, change the setup to
OFF
.
• Default temperature coefficient is
2.00%.
You can set a value in between
00.00% to
10.00%
.
• Default reference temperature is
25.0°C
.
You can set the value in between
15.0°C to
30.0°C.
• Erasing previous calibration data is recommended for accurate calibration. Default
calibration clear setup is
NO
but to erase the calibration data, change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip
Summary of Contents for LAQUA 2000 Series
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 54: ...Ion Setup 36 key key key key...
Page 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Page 98: ...Printout Summary 80 Salinity TDS...
Page 99: ...Printout Summary 81 Printer Format Data Log...
Page 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Page 101: ...Printout Summary 83 Conductivity Salinity...
Page 106: ...88...
Page 107: ......