Midas
®
Gas Detector
8-7
filter ensuring that the filter connectors are
fully engaged in the manifold ports and that
the filter is the correct orientation (arrow
pointing downwards). Orient the metal
retaining clamps away from the pump wiring
to prevent damage.
9. Route pump wiring away from the chassis and
manifold to prevent damage.
Diagram 8-7. Filter orientation.
Filter connector
Manifold ports
New filter
correct orientation
➭
8.5 System Leak Check Procedure
Caution
When performing a leak check, the Midas
®
must be placed in inhibit mode to prevent
false concentrations or faults to be activated
when plugging the Sample and/or Exhaust
Lines.
Example: O
2
(oxygen) – plugging of the ports
causes the O
2
levels within the Midas
®
flow
path to be depleted as the sample becomes
stagnant. The result is the concentration levels
begin to fall and trigger the alarms for the O
2
depletion.
1. Place the Midas
®
into Inhibit (see
Section
7.4.3
)
2. Plug the Inlet Sample line (see Diagram 8-8)
3. The flow meter indicators will begin to drop
and ultimately disappear (within seconds) (see
Diagram 2-3
)
4. Continue to keep the port plugged until
the Midas
®
reports an “F81” (Flow Fail) –
approximate time to fault is 1 minute
5. Remove the plug
6. Allow 15 seconds or so to allow the Midas
®
flow to stabilize then clear the fault by
pressing and holding ‘X’
7. Plug the Outlet Exhaust line
8. The flow meter indicators will begin to drop
and ultimately disappear (within seconds)
9. Continue to keep the port plugged until
Summary of Contents for Midas
Page 7: ...Midas Gas Detector 1 1 1 Description ...
Page 9: ...Midas Gas Detector 2 1 2 Overview ...
Page 14: ...Midas Gas Detector 3 1 3 Default Configuration ...
Page 17: ...Midas Gas Detector 4 1 4 Installation ...
Page 41: ...Midas Gas Detector 5 1 5 Startup ...
Page 43: ...Midas Gas Detector 6 1 6 Operation ...
Page 53: ...Midas Gas Detector 7 1 7 Navigating Modes and Submenus ...
Page 74: ...Midas Gas Detector 8 1 8 Maintenance ...
Page 82: ...Midas Gas Detector 9 1 9 Pyrolyzer Module Options ...
Page 89: ...Midas Gas Detector 10 1 10 Optional LonWorks Interface Installation ...
Page 99: ...Midas Gas Detector 11 1 11 Troubleshooting and Fault Diagnosis ...
Page 102: ...Midas Gas Detector 12 1 12 REFLEX ...
Page 104: ...Midas Gas Detector 13 1 13 Internal Web Server ...
Page 106: ...Midas Gas Detector 13 3 Diagram 13 1 IPAddress setting in Windows XP ...
Page 129: ...Midas Gas Detector 14 1 14 Typical Installation Topologies ...
Page 132: ...Midas Gas Detector 15 1 15 Ordering Information ...
Page 136: ...Midas Gas Detector 16 1 16 Specifications ...
Page 139: ...Midas Gas Detector 17 1 17 Calibration and Bump Testing ...
Page 143: ...Midas Gas Detector A 1 A Modbus TCP Interface ...
Page 150: ...Midas Gas Detector B 1 B Gas Tables ...
Page 158: ...Midas Gas Detector C 1 C Warranty ...