Midas
®
Gas Detector
7-21
7.4.4 Stimulate 4-20mA ‘4-20 mA’
1. Use the ‘
s
’ up or ‘
t
’ down buttons to select the
‘4-20’ submenu and press ‘
3
’ to accept.
2. The display shows ‘S 04’ indicating a 4 mA
output is being stimulated.
3. Use the ‘s’ up or ‘
t
’ down buttons to increase
or decrease current as desired in the range of
1 to 21 mA. Displayed as “S xx” = xx indicating
mA current being output.
4. Press ‘X’ to return to the ‘4-20’ submenu
selection.
5. Press ‘X’ to return to set-up, calibration and test
menu selection.
6. Press ‘X’ again to return to normal operation.
Caution
Proceeding with 20 mA stimulation may
cause unexpected alarm activation.
Only authorized operators should perform
this task!
This mode cancels itself in 5 minutes.
Summary of Contents for Midas
Page 7: ...Midas Gas Detector 1 1 1 Description ...
Page 9: ...Midas Gas Detector 2 1 2 Overview ...
Page 14: ...Midas Gas Detector 3 1 3 Default Configuration ...
Page 17: ...Midas Gas Detector 4 1 4 Installation ...
Page 41: ...Midas Gas Detector 5 1 5 Startup ...
Page 43: ...Midas Gas Detector 6 1 6 Operation ...
Page 53: ...Midas Gas Detector 7 1 7 Navigating Modes and Submenus ...
Page 74: ...Midas Gas Detector 8 1 8 Maintenance ...
Page 82: ...Midas Gas Detector 9 1 9 Pyrolyzer Module Options ...
Page 89: ...Midas Gas Detector 10 1 10 Optional LonWorks Interface Installation ...
Page 99: ...Midas Gas Detector 11 1 11 Troubleshooting and Fault Diagnosis ...
Page 102: ...Midas Gas Detector 12 1 12 REFLEX ...
Page 104: ...Midas Gas Detector 13 1 13 Internal Web Server ...
Page 106: ...Midas Gas Detector 13 3 Diagram 13 1 IPAddress setting in Windows XP ...
Page 129: ...Midas Gas Detector 14 1 14 Typical Installation Topologies ...
Page 132: ...Midas Gas Detector 15 1 15 Ordering Information ...
Page 136: ...Midas Gas Detector 16 1 16 Specifications ...
Page 139: ...Midas Gas Detector 17 1 17 Calibration and Bump Testing ...
Page 143: ...Midas Gas Detector A 1 A Modbus TCP Interface ...
Page 150: ...Midas Gas Detector B 1 B Gas Tables ...
Page 158: ...Midas Gas Detector C 1 C Warranty ...