description
maximum
Tubing OD, mm (in)
6.35 (0.25)
Tubing ID, mm (in)
4.76 (0.188)
Prepare Tubing
1. Cut the tube squarely and remove any burrs.
2. Mark from end of tube the length of insertion. The insertion lenght of Midas-Mfrom the end
of tube is 15.5 mm.
NOTE
When inserting the tube into the inlet/outlet port of the Midas-M, make sure that it is
inserted up to the marked position.
In-Line Filters
External filters must be used to protect the tubing from contamination. Use particulate filter part
number 780248 for normal gases and 1830-0055 or 1991-0147 for corrosive gases. Replace
the filter every 3 months. Refer to the Gases Table for specific gases.
Local Detector Option
The Midas-M gas detector can also be used to monitor for gas at the location of the detector. To
do this, an inline filter is simply connected to the sensor cartridge gas inlet port. The external
dust filter part number is 780248 for normal gases and 1830-0055 or 1991-0147 for corrosive
gases. The area around the detector is then being monitored as opposed to a sample being
drawn from a remote location.
Midas-M
18
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...