Test Mode
From the TestMode, you can test Bump, Simulation, 4-20mA, and Inhibit.
1. Go to the Monitoring Mode.
2. Press
or
buttons and hold for one second to enter the Main Menu, and one of the
Menu options is displayed.
3. Press
or
buttons to scroll to Test.
4. Press
to accept and enter the Test menu.
5. Press
or
to continue exploring more options from the Test Mode, and then press
to accep it.
6. Optional Step: Press
to cancel and return one step back.
From the Test Mode panel, you can test any of the following options:
Option
Operations you can Test from the selected option
Bump
Gas bump test
Simulation
l
Alarm 1 of each channel
l
Alarm 2 of each channel
l
Maintenance Fault
l
Instrument Fault
4-20mA
Force 4-20mA output of each channel
Inhibit
l
Inhibit ON/OFF
Midas-M
36
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...