CHAPTER
4
Maintenance
Midas-M is a fully serviceable product designed with modular components that can be readily
replaced by trained service personnel so as to minimize the time that the gas detector is not
available.
External in-line air filters should be replaced every three months or more frequently if the system
is sampling in environments that have high levels of particulate matter or very acidic / wet
atmospheres. Similarly, the internal particulate filter should be replaced every two years or more
frequently if the sample lines are prone to heavy contamination. Refer to the following table.
Every sensor cartridge is shipped with a 2 years warranty. All sensor cartridges are factory
calibrated to traceable national standards before shipment to the end user.
Note that testing or calibrating with the wrong (incorrect, out of date, non-traceable) calibration
gases, calibration equipment, methods or operating conditions can actually damage the sensor
cartridge’s lifetime and alter the calibration adversely. Only qualified calibration technicians
should attempt to calibrate the Midas-M gas detector.
The internal pump module is designed to operate for a minimum of 24 months and it is
recommended that this pump module (part number MM-PM ) be replaced every 2 years.
Component
Frequency
Pump
2 years or as needed
Internal filter
2 years or as needed
External Sample Line Filter
Part Number 780248, every 3-6 months
Part Number 1991-0147, every 3-6 months
Part Number 1830-0055, every 3-6 months
Leak Check
Leak Check every 6 months or after replacing any component.
Bump Test
6 months
Flow Calibration
Flow Calibrate after pump, internal and external filter replacement.
Recommended maintenance schedule
Midas-M
65
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...