OC-905

Portable gas detector

Operating Manual

Henan Oceanus

Summary of Contents for OC-905

Page 1: ...OC 905 Portable gas detector Operating Manual Henan Oceanus...

Page 2: ...ge accessories arbitrarily and does not conform to the above mentioned manual the operator shall be responsible for the reliability of the instrument And this company is not responsible for the warran...

Page 3: ...adopts the full metal shell OC 905 adopts the brand new imported sensor and 32 bit low power processor the circuit adopts four wiring which has a good effect on anti interference and week signals the...

Page 4: ...to disassemble this instrument voluntarily in order to guarantee that the using of the detector is safe and reliable Please do not put this detector in the high temperature moist or strong static envi...

Page 5: ...ept coal mine and GB50257 1996 the relevant regulations of the explosion of the installation works of electrical equipments and the construction of electrical installation in the fire dangerous enviro...

Page 6: ...press shutdown key In the measuring process if some parameters of the machine need to be set enter into the menu interface to set The specific operation process is as follow 3 1 Startup In shutdown mo...

Page 7: ...ry capacity b In the second line co2 is the gas type and the machine display the different name for the different gas types c In the third line the big character 850 6 is the air concentration value t...

Page 8: ...h has the storage data function g 7 RESTORE SET restore factory setting the parameter setting of factory can be restored h 8 ABOUT OC 905 some information which is related to this machine such as vers...

Page 9: ...erface Note the one level alarm value cannot be bigger than the second level value and the alarm value cannot be bigger than the scale span value 3 5 2 Time and date setting On the menu interface pick...

Page 10: ...ter picking on the calibration setting menu the system would enter into the password interface firstly and put into password 88 and then enter into the calibration interface On the calibration interfa...

Page 11: ...alibration interface make the zero cleaning operation This operation is made on the situation that there is no object gas 2 The span calibration needs the participation of the object gas if customer d...

Page 12: ...d 10 minutes After choosing the corresponding storage time internal press confirmation key and open the storage function If the measuring is completed choose OFF again and then press confirmation or O...

Page 13: ...uccessfully the interface would display MKFS OK or would display MKFS ERR Choose BACK to return and also can press return key to return Note 1 This function is only valid to the machine which has the...

Page 14: ...V Basic calibration The product has been calibrated before leaving factory If the user has the calibration conditions they can make the periodic calibration according to the manual in the process of u...

Page 15: ...g principle Detecting principle Electrochemistry infrared detection PID thermal conductivity catalytic combustion Display mode 2 5 inch Lattice LCD liquid crystal digital display Backlight Highlight w...

Page 16: ...12 hours The continuous working 12 hours normal temperature electrochemical sensor Power consumption 0 5W Working pressure 86kPa 106kPa Weight About 0 46 Kg Explosion proof sign Exd ib IIC T4 Gb Prot...