Specifications
Specifications
Transmitter Dimension
Size (unit with
Cartridge)
"136 mm (H) X 83 mm (W) X 152 mm (D) (5.35 X 3.27 X 5.98 in)"
Weight -
Transmitter
1.3 kg (2.87 lb)
Weight -
Cartridge
0.17 ~ 0.22 kg (0.38 ~ 0.49 lb) dependent on sensor type
Optional Relay Dimension
Size
"137 mm (H) X 84 mm (W) X 41 mm (D) (5.39 X 3.31 X 1.61 in)"
Weight
0.31 kg (0.68 lb)
Power Requirements
Operating
Voltage
24 VDC Nominal, -15 to +10% (20.4 to 26.4 VDC)
"Operating
Voltage with
Power over
Ethernet (PoE)"
48 VDC PoE (IEEE 802.3af compliant)
Power Consumption
Transmitter unit
(normal
condition)
1
Typ. 5W
Transmitter unit
(full load
condition)
2
≤ 11.45W
Transmitter with
Optional Relay
≤ 12.9 W
Outputs
Visuals
"Alarm, power, fault LEDs, and LCD with all the gas readings and
events. LEDs: Power (Green), Alarm 1 (Red), Alarm 2 (Red), Fault
(Yellow)"
Relays in
Transmitter
"Alarm1, Alarm2, Fault Relays (3) rated 1.0 A @ 30Vdc or 0.5A @
125Vac Max 10uA @ 10mV minimum, configurable as normally open
or closed, latched or unlatched"
Midas-M
74
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...