ModBus
Register
Number
Information R/W
Data
Type
Data
Size
(Byte)
Description
Note
40027 to
40028
Device
Instrument
Fault
R
u32
4
This is the bit
order
representation
of the
instrument
fault status. If
any fault exists
this will take a
value in each
bit. if any
instrument
fault exists, this
will take a
value zero.
The meaning
of bit will be
described as
fault code in
more detail
bit 0 : LCD Fault (F01) bit 1 :
Simulated Instrument Fault (F02) bit 2
: Reserved for future expansion bit 3 :
Flow Fail (F41) bit 4 : Pump EEPROM
(F42) bit 5 : Pump Absent (F43) bit 6 :
LIT Fail (F44) (Reserved) bit 7 :
Cartridge Absent (F60) bit 8 : Catridge
Wrong Type (F61) bit 9 : Cartridge No
valid cell (F62) bit 10 : Catridge
EEPROM (F63) bit 11 : Catridge
Flash (F64) bit 12 : Catridge Power
3.3v (F65) bit 13 : Catridge Power 5v
(F66) bit 14 : Catridge Application Fail
(F67) bit 15 : Transmitter Nor Flash
(F80) bit 16 : Transmitter EEPROM
Corrupted (F81) bit 17 : Transmitter
EEPROM Com Fail (F82) bit 18 :
Transmitter Flash (F83) bit 19 :
Transmitter Power 3.3v (F84) bit 20 :
Transmitter Power 24v (F85) bit 21 :
Transmitter Power 5v (F86) bit 22~31
: for future expansion
40029
Relay Status
R
u16
2
The meaning
of bit : Value 0
is De-
energized
Value 1 is
Energized
bit 0 : Internal Relay 1 energized bit 1 :
Internal Relay 2 energized bit 2 :
Internal Relay 3 energized bit 3 : for
future expansion bit 4 : External Relay
Fitted bit 5 : External Relay 1
energized (Ch1 alarm1) bit 6 :
External Relay 2 energized (Ch1
alarm2) bit 7 : External Relay 3
energized (Ch2 alarm1) bit 8 :
External Relay 4 energized (Ch2
alarm2) bit 9 : External Relay 5
energized (Any fault) bit 10 : External
Relay 6 energized (Ch3 alarm1) bit 11
: External Relay 7 energized (Ch3
alarm2) bit 12 : External Relay 8
energized (Ch4 alarm1) bit 13 :
External Relay 9 energized (Ch4
alarm2) bit 14~15 : for future
expansion
40030
Temperature
in Celsius (oC)
R
s16
2
Temperature
Midas-M
90
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...