ModBus
Register
Number
Information R/W
Data
Type
Data
Size
(Byte)
Description
Note
40023
Decimal Point
Indicator
R
u16
2
Decimal point
indicator of
Display Gas
Concentration
bit 0 ~ 2 : Ch1 Decimal point indicator
(0, 1, 2, 3) bit 3 ~ 5 : Ch2 Decimal
point indicator (0, 1, 2, 3) bit 6 ~ 8 :
Ch3 Decimal point indicator (0, 1, 2, 3)
bit 9 ~ 11 : Ch4 Decimal point
indicator (0, 1, 2, 3) bit 12~15 : for
future expansion
40024
Heat beat
Counter
R
u16
2
System time in
seconds.
Range of this
value is from 0
to 59
40025 to
40026
Device
Maintenance
Fault
R
u32
4
This is the bit
order
representation
of the
maintenance
fault status. If
any fault exists
this will take a
value in each
bit. if any
maintenance
fault exists, this
will take a
value zero.
The meaning
of bit will be
described as
fault code in
more detail
bit 0 : Simulated Maintenance Fault
(m02) bit 1 : Flow Warning (m41) bit
2~31 : for future expansion
Midas-M
89
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...