7.2.4 Setting the network parameters
i
For parameterization of the EK280 you need a connection cable for the optical interface as well
as the enSuite program. enSuite can be downloaded free of charge from the Honeywell website
in the "Software Downloads" section:
www.elster-instromet.com/en/software-downloads
a) Closing the time windows
Time windows 1 to 4 must be closed for parameterization to function securely. To do this:
−
Connect a PC or laptop with enSuite installed to the EK280 optical interface using the connection
cable.
−
Start enSuite and establish the communication link to the EK280.
(
3.1 Setting up the communication link to the EK280 via optical interface
−
Select “
Navigation
” on the left-hand side and then “
Parameterize online
” at the bottom of the
navigation window.
−
In the top part of the device window, open the path “
Parameters
” > “
Interfaces
” by clicking on the
relevant plus signs, and then select “
Call time windows
”.
−
In call windows 1 to 4, set start and end to the same value for each one and then select the “
Transfer
changes to device
” icon at the top of the enSuite window.
−
Then wait a minute for any communications activity in progress on the EK280 to be safely stopped.