Press the ENTER key to confirm the set value. It is possible to abort the entry by pressing the ESC
key.
9.8 Calibration data log
With the aid of the calibration data log, several parameters which are relevant from a statutory calibration
viewpoint (
) can be changed when the administrator lock is open.
A data row is entered for each alteration of such a parameter when the calibration lock is closed. In addition,
an entry is always made when opening and closing the calibration lock.
i
The calibration data log is not a flexible archive.
9.8.1 Deactivating the calibration data log
−
In order to deactivate the calibration data log on the device, open the calibration lock as described in
the
−
Move the cursor to
"Serv."
>
"Archives" > "CDL".
−
Press the ENTER key.
⇒
"on"
flashes.
−
Change the value using the arrow keys
or
to
"off"
.
−
Press the ENTER key to confirm the set value. It is possible to abort the entry by pressing the ESC
key.
i
If the calibration data log is deactivated then the values affected are locked in a calibration lock.