Marker Data (Spot Noise)
The following commands allow users to read a single data point from the instrument at any
offset frequency. The instrument will return the data point at the offset frequency nearest
the specified value by default. Linear interpolation may be enabled, and then instrument will
interpolate a data point at the exact specified frequency offset value.
Read Marker Data (Spot Noise)
:CALC:PN:TRAC:MARK?<value>
Returns the phase noise value at specified
frequency offset
Example
TX:
:CALC:PN:TRAC:MARK?10kHz
RX:
10.003, -1.2177e+02
<value> = frequency offset (in Hz unless kHz, MHz are specified)
NOTE:
The response is formatted as (freq offset, amplitude)
Marker Data Interpolation
:CALC:PN:TRAC:MARK:INTERP<value>
Enable/disable interpolation
Example
TX:
:CALC:PN:TRAC:INTERP:ON
RX:
Marker interpolation on
<value> ON, OFF
:CALC:PN:TRAC:MARK:INTERP?
Query interpolation status
Example
TX:
:CALC:PN:TRAC:INTERP?
RX:
Marker interpolation ON