HA7062D Manual Rev 1.01
www.holzworth.com
Page 85 of 105
HX5100 Phase Shifters
:SENS:PN:HX5100:<value>
Tells the instrument if HX5100 phase shifters
are in use (1 = in use, 0 = not in use)
Example
TX:
:SENS:PN:HX5100:1
RX:
HX5100 in use
<value> 0, 1
:SENS:PN:HX5100?
Queries HX5100 setting
Example
TX:
:SENS:PN:HX5100?
RX:
HX5100 in use
Load HX5100 Calibration Data
(
each calibration file line must be loaded individually in the order it appears in the file
)
:MMEM:PN:CAL:HX5100:CHn:DATA:<value>
Loads
one line
from HX5100 calibration file
(
n
=
Channel #)
Example
TX:
:MMEM:PN:CAL:HX5100:CH2:DATA:<value>
RX:
HX5100 Channel 2 calibration data processed
<value> See next page
:MMEM:PN:CAL:HX5100:VALID?
Check if all calibration data was loaded
successfully
Example
TX:
:MMEM:PN:CAL:HX5100:VALID?
RX:
"HX5100 cal data valid" OR "HX5100 cal data invalid. Missing frequency
data" OR "HX5100 cal data invalid. Missing voltage data" OR "HX5100
cal data invalid. Missing delay data for channel 1" OR "HX5100 cal data
invalid. Missing delay data for channel 2"