2.
If the instrument is not present in Device Manager or in the Holzworth application GUI please
unplug the USB cable and power cycle the analyzer. Wait 5-10 seconds for the analyzer to initialize
an re-insert the USB cable. Click
Locate Devices.
3.
If the analyzer is still not detected download the device drivers may need to be manually
installed. Download and extract the executable using the link below. Run the executable to ensure
the proper device drivers are installed.
Device drivers:
http://www.holzworth.com/software/Analyzers/HA7000D/CDM21228_Setup.zip
After the executable has finished installing the drivers repeat troubleshooting Step 1.
4.
Attempt to make a connection through a 'USB hub' if available. Upon connecting through a hub it
may be necessary to repeat troubleshooting Step 1.
5.
Contact Holzworth Support for further assistance.
9.2.2 GPIB COMMUNICATION
The Holzworth HA7062D is GPIB capable. ASCII commands to configure GPIB settings are listed
in Appendix C. These commands may be issued using the Console via a USB or Ethernet
connection. After configuring the GPIB settings the instrument can be connected to the bus.
STEP THREE
Under Ports (COM & LPT)
locate COM port associated
with the HA7062C (identified as
"USB Serial Port")