HA7062D Manual Rev 1.01
www.holzworth.com
Page 89 of 105
Calibrate LO1 and LO2 (Internal and External LO's)
:INIT:PN:CAL
Begins the calibration process for LO1 and
LO2
Example
TX:
:INIT:PN:CAL
RX:
LO Calibration started
Instrument Status
:STAT:OPER:COND?
Returns the status of the instrument
Example
TX:
:STAT:OPER:COND?
RX:
"Instrument Busy" OR "Instrument Ready"
NOTE:
Returns "Instrument Busy" during calibration, returns "Instrument Ready" when calibration is
complete. Continuously sending this command during calibration will slow the process.
Cancel Calibration Routine
:CALC:PN:TRACE:HOLD
Stops the LO calibration routine
Example
TX:
:CALC:PN:TRACE:HOLD
RX:
Calibration canceled