11.3 SINGLE CH. ADDITIVE MEASUREMENT WITH EXTERNAL MIXER
This example utilizes a 3GHz source, however using the instrument to perform an additive
measurement in this manner will allow the user to measure the phase noise at frequencies >6GHz.
Below is a detailed block diagram for performing a single channel additive phase noise
measurement using an external mixer with the HA7062D Phase Noise Analyzer. External Additive
mode utilizes the Baseband inputs for direct access to the FFT engine of the HA7062D.