9.9.3 LO SOURCE OUTPUTS
LO Source Outputs
provides the user access to the frequency and power settings on the HSX
series synthesizers which function as internal LO's for the HA7062D analyzer.
9.9.4 LO TUNE OUTPUTS
LO1/LO2 Tune Voltage:
Sets instantaneous voltage level at tune voltage outputs.
LO1/LO2 Start/Stop:
Sets the tuning voltage range for External LO's. Settable from -10V to
+12V. Accurately setting the tune voltage range of each External
LO is necessary for achieving
a valid calibration lookup table and phase locking. For external LO mode the phase noise
performance of each external LO should be similar, however the tune voltage slope of the LO's
need not be similar. Furthermore, their tune voltage slopes could be completely opposite one
another.