2.6.3 Operation test of viscometer
-
This function allows you to check the viscometer operation and measure the current ink
viscosity. Note that this operation test is not to determine the coefficient.
-
It is not necessary to use new ink for operation test.
1 Press Viscometer calibration on page 2 of circulation control of service
maintenance screen.
2 Press Operation test.
3 When the test is complete, the current “ink viscosity” will appear.
If the process status is “fault” and message “Check viscometer for any defect”
appears, normal measurement has not been executed. Follow “4.9 Troubleshooting
when fault occurs in viscometer”.
4 Press Cancel to restore the original screen.
The operation test result display will disappear.
Cancel
Plunger falling time
No.1 X. XX (sec)
Standard X. XXX (mPa.s)
Coefficient XXXX
Average X. XX (sec)
No.2 X. XX (sec)
No.3 X. XX (sec)
No.4 X. XX (sec)
No.5 X. XX (sec)
Ink temperature XX (
o
C)
<the previous calibration>
Exec.time 2011.07.07 12:49
Status:XXXXXXXXX
2011.07.08 09:12
Viscometer calibration
Plunger falling time
No,1 X. XX (sec)
Standard X. XXX (mPa.s)
Coefficient XXXX
Average X. XX (sec)
No.2 X. XX (sec)
No.3 X. XX (sec)
No.4 X. XX (sec)
No.5 X. XX (sec)
Ink temperature XX (
o
C)
<the current calibration >
Exec.time 2011.07.08 09:12
Proc. status: Ended
Ink viscosity XXX
It is not necessary to calibrate the viscometer
for PXR-H and PXR-P, therefore “the previous
calibration” is not displaed.
Summary of Contents for PXR
Page 1: ...Service Manual HITACHI Printer Model PXR Revision December 2008 Version C ...
Page 80: ...3 17 3 17 Circuit diagram of EZJ95 ...
Page 215: ...8 2 8 1 2 Circulation system diagram PXR P ...
Page 216: ...8 2 Air system diagram PXR P 8 3 ...
Page 217: ...8 4 8 3 Electrical Connection Diagram 8 3 1 Electrical Connection Diagram PXR D ...
Page 218: ...8 5 8 3 2 Electrical Connection Diagram PXR H ...
Page 219: ...8 3 3 Electrical connection diagram PXR P 8 6 ...