54
Function Specifications
4.3
Function Specifications
Key input
keys” (p. 16), and “Model L9788-10 Test Lead with
Auto power save
(APS)
Automatically turns off the power 15 minutes±1
minutes after the last measurement (measurement
voltage is 5 V or less for Voc mode) with the power on.
Recovering from auto power save:
Set the rotary switch in OFF to turn off the power and
then turn on the power once again.
Indication
Semi-transmissive FSTN LCD, positive
Battery indicator
Battery voltage
Battery level
7.95 V±0.2 V or higher
3
7.35 V±0.2 V to 7.95 V±0.2 V
2
6.75 V±0.2 V to 7.35 V±0.2 V
1
6.75 V±0.2 V or less
0 (Battery
shape flashes)
• Hysteresis
Low battery level remains even if battery voltage
increases but the power is not turned on once
again.
• Measurements cannot be carried out when battery
level is 0.
(
[bAt Lo]
is displayed and automatically turns off
the power.)