51
Input/Output/Measurement Specifications
Duration of
shorting between
terminals
10 ms or less
Output pulse
Voltage: 100 V DC or less
Pulse width: 5 ms or less
Limiting current: Measured short-circuit c 1 A or
less
Maximum: 13 A
Judgment
method
Detected peak current ≥ Measured short-circuit current
+ additional current × 0.8: PASS
Detected peak current < Measured short-circuit current
+ additional current × 0.8: FAIL
R
BPR
condition can be added to the conditions above
with comparator setting.
Measurement
time
2 seconds or less (3 seconds or less when
measurement voltage is 10 V or less)
Current
measurement
method
Peak value detection
Measuring range
for bypass route
resistance
0
Ω
to 15
Ω
The R
BPR
value will flash if insolation (short-circuit
current) varies during measurement.
(3) COUNT mode
Measurement
item
Displays the number of bypass diode measurements
Number of
measurements
Counts the number of measurements from power on to
power off.
Resets the count to 0 when the power is turned off or
auto power save.
Maximum count: 199
More than 200 counts:
[>199 V]
flashes
4
1
2
3
5
6
7
Appx.
Index