31
Pass/Fail Judgment (Comparator) Function
3.4 Pass/Fail Judgment (Comparator)
Function
This function compares the measured value with the preset value (= judgment
reference value) and judges whether the result is PASS (good) or FAIL
(defective).
Comparator settings for each mode pass/fail judgment (comparator) will be
saved, even if the rotary switch is set in OFF.
The range that can be set for the reference value is as follows:
DIFF
0 V to ±20 V. The setting allows 1 V increments.
R
BPR
0
Ω
to 15
Ω
. The setting allows 0.5
Ω
increments.
Indication lights up
PASS (good) judgment
FAIL (defective) judgment
Indication
Backlight:
No change
Lights up in
green
Backlight:
Lights up in red
Lights up in
red
Type of measurements that can be judged
Mode
PASS judgment
FAIL judgment
State of measured value
Buzzer Backlight Buzzer
Voc
Reference value or lower
Short
beep
Lights up
in red
Long
beep
BPD TEST
Lower than reference value
and bypass diode is working
correctly
1
2
4
5
6
7
Appx.
Index
3
Summary of Contents for FT4310