36
Making Measurements in Voc Mode
Auto-hold
With the comparison display function, when the measured value
stabilizes, a short beep is emitted and the measured value is
automatically retained.
When the measured value is 20 V or higher and stabilizes for
approx. 1 second (fluctuation range is within 20 V), the measured
value is automatically retained.
When the test leads are disconnected from the measurement object
and a new measurement is started, the new measured value is
retained.
[HOLD]
flashes: Waiting for stabilization of the measured value
[HOLD]
appears: The measured value is retained
Conceptual diagram of auto-hold (AC voltage)
Indicated value
Auto-hold
Example: 100 V
Auto-hold
Example: 99 V
Threshold allowing
auto-hold
100 V
99 V
Measured
value
Connect to the
measurement
object
Disconnect Connect Disconnect Connect
Disconnect
Dead zone
Auto-hold is
not applied
(threshold is
not exceeded)
Stabilized
range