24
Using Model L9788-10 Test Lead with Remote Switch (Red)
2.2 Using Model L9788-10 Test Lead
with Remote Switch (Red)
Pre-measurement Inspection
1
Set the rotary
switch in OFF.
2
Fully insert the plug of L9788-10 into
terminal
P(+)
of the instrument.
OFF
OK
NO
3
Set the rotary switch in SELF CHECK.
4
Press the
RELEASE
key to cancel the lock function.
5
After
[Short the Test Leads]
is shown on the display, short
the test lead tips with each other.
6
Press the
MEASURE
key of the L9788-10.
Verify that the display indicates
[PASS]
.
Lights up in green
PASS
7
Press
.
Check that the L9788-10 tip
lamp lights up.