Hioki FT4310 Instruction Manual Download Page 32

28

Pre-measurement Inspection

From the previous page

Checking the remaining battery charge

Is the battery level adequate?

Set the rotary switch in SELF 

CHECK and confirm the 

battery indicator.

Flashes

Replace the batteries. (p. 22)

 

  appears

Inspection complete

Please read “Operating Precautions” (p. 8) before use.

Summary of Contents for FT4310

Page 1: ......

Page 2: ......

Page 3: ...itch Red 24 2 3 How to Use the Carrying Case Model C0206 26 3 Measurement 27 3 1 Pre measurement Inspection 27 3 2 Auto Power Save Power Saving Function 29 3 3 Auto Backlight off Automatic Light off F...

Page 4: ...aking Measurements with the Bluetooth Function 46 4 Specifications 47 4 1 General Specifications 47 4 2 Input Output Measurement Specifications 49 4 3 Function Specifications 54 4 4 Interface Specific...

Page 5: ...rred during shipping In particular check the accessories panel switches and connectors If damage is evident or if it fails to operate according to the specifications contact your authorized Hioki dist...

Page 6: ...vailable for the instrument Contact your authorized Hioki distributor or reseller when ordering 1 2 Product name Maximum rated voltage and maximum rated current 1 Model L9788 92 Breaker Pin CAT III 60...

Page 7: ...damage to the instrument Be certain that you understand the instructions and precautions in the manual before use WARNING Protective gear This instrument is measured on a live line To prevent an elect...

Page 8: ...en displays are enclosed in square brackets MEASURE Bold Indicates the names of keys and other controls on the instrument Symbols on the instrument Indicates cautions and hazards When the symbol is pr...

Page 9: ...isplays the alphanumeric characters as follows 1 2 3 4 5 6 7 8 9 0 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z Accuracy We define measurement tolerances in terms of rdg reading and dgt digt va...

Page 10: ...trademark is used by HIOKI E E CORPORATION under license Android Google Play are registered trademarks of Google Inc IOS is a registered trademark of Cisco in the U S and other countries iPhone iPad i...

Page 11: ...g so could result in a serious accident CAT II When directly measuring the electrical outlet receptacles of the primary electrical circuits in equipment connected to an AC electrical outlet by a power...

Page 12: ...cting test lead tip To prevent an electric shock be careful to avoid shorting live lines with the test leads tip If the test lead or the instrument is damaged there is a risk of an electric shock Perf...

Page 13: ...h Do not use the instrument with circuits that exceed its ratings or specifications Doing so may damage the instrument resulting in an electric shock Use only the specified test leads Use of any test...

Page 14: ...erator Maximum rated voltage to terminal of the instrument is 1000 V DC Do not use the instrument for equipment with rated voltage over 1000 V DC Doing so may cause an electric shock or damage to the...

Page 15: ...xposed to a strong electromagnetic field or electrostatic charge Near induction heating systems such as high frequency induction heating systems and IH cooking equipment Susceptible to vibration Expos...

Page 16: ...tal pins at the end of the test leads To prevent a short circuit accident be sure to use the test leads with the sleeves attached when performing measurements in the CAT III measurement category Remov...

Page 17: ...surements Bypass diode condition can be measured with a connection box Measurement results can be sent to a tablet or other mobile device by Bluetooth Detecting bypass diode deterioration The resistan...

Page 18: ...ment switch 2 N terminal Connect the black test lead 3 Control terminal Connect Model L9788 10 Test Lead with Remote Switch Red 4 P terminal 5 Rotary switch p 15 Selects the measurement functions the...

Page 19: ...Setting the rotary switch in a position other than OFF turns on the power Rotary switch state Voc mode BPD TEST mode Self check mode Description in this manual Set the rotary switch in Voc Set the rot...

Page 20: ...ode screen When comparator is set Decrease of judgment reference value p 32 When comparison display is set Reference value reset p 35 COUNT mode screen Count reset p 51 3 COMP Bluetooth In Voc or BPD...

Page 21: ...s when the measured value is less than the minimum display value Flashes when the measured value is greater than the maximum display value Appears when the measured value is retained p 36 Appears when...

Page 22: ...s measurement p 15 p 35 Appears when the comparator function is enabled p 31 Reference value for comparison display measurement in Voc mode p 35 Difference between reference value and measured value f...

Page 23: ...Plug Enlarged tip view 35 16 1 8 3 0 Unit mm 1 Light Lights up interlocked with the backlight of the instrument 2 MEASURE key Measurement switch Press Press the MEASURE key Press and hold Turns on th...

Page 24: ...20 Names and Functions of Parts...

Page 25: ...1 2 Preparing for Measurement 1 Insert the batteries p 22 2 Connect the test leads to the terminals N terminal Connect the black test lead P terminal Connect the red test lead 1 3 4 5 6 7 Appx Index 2...

Page 26: ...electric shock use only the screw for securing the battery cover in place that are originally installed If you have lost a screw or find that a screw is damaged please contact your authorized Hioki d...

Page 27: ...teries in accordance with local regulations OFF 1 Set the rotary switch in OFF Disconnect the test leads from the instrument Battery cover Screw 1 Tabs Battery LR6 6 Check the polarity 2 Loosen the fa...

Page 28: ...g of L9788 10 into terminal P of the instrument OFF OK NO 3 Set the rotary switch in SELF CHECK 4 Press the RELEASE key to cancel the lock function 5 After Short the Test Leads is shown on the display...

Page 29: ...ary switch in OFF and disconnect the L9788 10 Model L9788 90 Tip Pin Wrench size 7mm width 2 Remove the tip pin by rotating with a wrench 3 Attach the new tip pin to the L9788 10 by rotating with a wr...

Page 30: ...the sleeves and firmly push them all the way in 2 3 How to Use the Carrying Case Model C0206 CAUTION Attach the strap securely to the four fittings on the carrying case If insecurely attached the ins...

Page 31: ...o not use any damaged cables and replace them with those specified by our company This may cause an electric shock Not Exposed Checking the instrument The following issues may be occurring The test le...

Page 32: ...e Checking the remaining battery charge Is the battery level adequate Set the rotary switch in SELF CHECK and confirm the battery indicator Flashes Replace the batteries p 22 appears Inspection comple...

Page 33: ...15 minutes after the last operation or live wire warning indication After use set the rotary switch in OFF The auto power save feature consumes a small amount of current Disabling the function Other...

Page 34: ...3 minutes has passed since the last operation The automatic light off function can be canceled when working continuously in a dark location Disabling the function Continuous short beep Press for appro...

Page 35: ...or the reference value is as follows DIFF 0 V to 20 V The setting allows 1 V increments RBPR 0 to 15 The setting allows 0 5 increments Indication lights up PASS good judgment FAIL defective judgment I...

Page 36: ...COMP flashes and a value set as judgment reference value at that moment is displayed 3 3 Press or to set the reference value If you do nothing about 2 seconds after you set the reference value the com...

Page 37: ...Disconnect lightning arresters from the measuring circuit Disconnection is not required for the figure solar generator facility shown below because a lightning arrester is not present at the string si...

Page 38: ...hort circuit fault WARNING During measurement do not switch over to other measurement modes N P 1 2 3 4 Power Conditioner Main switch String N P 1 Set the rotary switch in Voc 2 Connect the black test...

Page 39: ...nd CONT appears HOLD appears Example measurement results display 2 Connect the test lead to the measurement object HOLD appears and Voc and REF show their measured values Example measurement results d...

Page 40: ...test leads are disconnected from the measurement object and a new measurement is started the new measured value is retained HOLD flashes Waiting for stabilization of the measured value HOLD appears T...

Page 41: ...g the open circuit voltage of solar battery string When the voltage is 1V or lower red and white of the backlight flash alternately Ground Fault Detection Connecting the black test lead to the ground...

Page 42: ...asure under a live circuit condition Doing so could damage the instrument or cause an accident that might result in injury or death Always turn off power to the measurement object before starting When...

Page 43: ...even if the rotary switch is set in BPD TEST and the MEASURE key is pressed Canceling the lock function Flashes 1 1 Set the rotary switch in BPD TEST RELEASE key flashes Off 2 2 Press Pressing this ke...

Page 44: ...annot be judged correctly N P 1Release 2 1 Release the MEASURE key 2 Set the rotary switch in BPD TEST Press to cancel the lock function p 39 N P Power conditioner Main switch String 3 4 OFF N P 3 Con...

Page 45: ...tion is engaged no measurement starts Verify the test lead connections once again If the solar panel is not generating power due to the shielding of the light continuous measurement and negative volta...

Page 46: ...ents in BPD TEST Mode After measurements are completed 1 Reconnect the lightning arrester connection if disconnected 2 Turn on all the disconnector devices of the strings 3 Turn on the main switch of...

Page 47: ...s iPhone iPad iPad mini iPod Touch and Android devices For more information about this functionality see the help function in the application software GENNECT Cross 1 Install the GENNECT Cross on your...

Page 48: ...or other penalties as a violation of applicable laws or regulations For more information see the attached Precautions Concerning Use of Equipment that Emits Radio Waves or go to our website The avail...

Page 49: ...connection setup screen simply move it close to the instrument to automatically pair it with the instrument the application software can be paired with up to 8 instruments Allow about 5 to 30 seconds...

Page 50: ...nt logging or waveform display to start a measurement For more information about each function see the help function in the GENNECT Cross The values displayed by the instrument may be different from t...

Page 51: ...or less no condensation 50 C 122 F or higher but less than 55 C 131 F 40 RH or less no condensation 55 C 131 F or higher but less than 60 C 140 F 30 RH or less no condensation 60 C 140 F to 65 C 149 F...

Page 52: ...at 23 C as a referential Approx 3000 times Comparator LCD backlight Bluetooth Off Interface Bluetooth Dimensions Approx 152W 92H 69D mm 5 98 W 3 62 H 2 72 D Mass Approx 650 g 22 9 oz including batter...

Page 53: ...BPD TEST Input terminal P terminal Connected to positive terminal of PV string N terminal Connected to negative terminal of PV string Maximum input voltage 1000 V DC Maximum rated voltage to earth 60...

Page 54: ...Measurement voltage 20 V or higher Fluctuation range of measurement voltage for 1 second 20 V or less Auto hold update A signal meeting the above Auto hold conditions is input after a measurement volt...

Page 55: ...o the conditions above with comparator setting Measurement time 2 seconds or less 3 seconds or less when measurement voltage is 10 V or less Current measurement method Peak value detection Measuring r...

Page 56: ...uaranteed accuracy period 1 year Guaranteed accuracy period from adjustment made by Hioki 1 year Temperature and humidity for guaranteed accuracy 23 C 5 C 73 F 9 F 80 RH or less Power voltage range wi...

Page 57: ...ter than 13 0 A but less than or equal to 15 0 A Overrange display 15 0 A or 0 0 A is displayed 3 Bypass route resistance measurement Range displayed range 15 0 0 to 15 0 Accuracy range 0 0 to 15 0 Ac...

Page 58: ...Set the rotary switch in OFF to turn off the power and then turn on the power once again Indication Semi transmissive FSTN LCD positive Refer to Display p 17 Battery indicator Battery voltage Battery...

Page 59: ...falls to 5 V or less or after the last operation Automatic polarity judgment function When measurement voltage is 1 V or less Buzzer sounds red backlight flashes and white backlight lights up Auto hol...

Page 60: ...ff Threshold value PASS BPD TEST mode Threshold value can be set in 0 5 increments from 0 to 15 for RBPR RBPR Threshold value FAIL RBPR Threshold value and Detected peak currentd Measured short circui...

Page 61: ...c Attribute Profile Supported Android devices Android4 3 or later Bluetooth low energy enabled devices Supported iOS devices iOS10 or later Bluetooth low energy enabled devices 4 5 Factory Settings Se...

Page 62: ...58 Factory Settings...

Page 63: ...ions when transporting the instrument When transporting the instrument for repair or any other purpose be sure to observe the following precautions To avoid damage to the instrument remove the batteri...

Page 64: ...batteries Replace the batteries with new alkaline batteries p 22 Has 1 minute lapsed after the last operation Release the lock function again p 39 The MEASURE key of test lead with remote switch is no...

Page 65: ...ssary Contact your authorized Hioki distributor or reseller Error indication Description Remedy and Reference Zero adjustment error resistance value of 3 or greater during the self check Check whether...

Page 66: ...is required Bluetooth is broken Repair is required AC voltage input Verify that you have connected the instrument properly Low battery voltage Replace the batteries p 22 1 V or greater input during t...

Page 67: ...urrent IB Short circuit current IA 1 A current If short circuit current IA applied current IB the bypass diode is working correctly OK If short circuit current IA applied current IB the bypass diode h...

Page 68: ...voltage of another solar panel measurement object voltage If reference voltage VA measurement object voltage VB the bypass diode is working correctly OK If reference voltage VA measurement object volt...