4.1 Comparator Function
53
4
C
ha
pte
r 4
A
pp
lie
d M
ea
su
re
m
en
t
Two methods are available for setting comparator thresholds.
Comparator Threshold Method Selection
1
(The SHIFT indicator lights up.)
The Comparator setting display appears.
2
Press so that the indicated position blinks, and set the comparator threshold
method.
HIGH, LOW
..... Compare against specified upper and lower thresholds (default setting
method)
REF, %
............ Compare against upper and lower thresholds internally calculated from a
specified reference value and tolerance
About comparisons based on a reference value and tolerance
When the reference value and tolerance method is selected, thresholds are calculated as
follows:
Upper threshold = reference value
×
(100 + tolerance [%]) / 100
Lower threshold = reference value
×
(100 - tolerance [%]) / 100
Measured values are displayed as a percentage relative to the reference value, calcu
-
lated as follows:
Relative value = (measured value - reference value) / reference value
×
100 [%]
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...