8.6 Message Reference
136
Set and Query the Comparator Upper Threshold Setting
(Resistance Measurement)
Syntax
Command
:CALCulate:LIMit:RESistance:UPPer
<Upper threshold>
Query
:CALCulate:LIMit:RESistance:UPPer?
Response
<Upper threshold>
<Upper threshold>
= 0 to 99999
(NR1)
Example
Command
:CALC:LIM:RES:UPP 28593
Sets the upper threshold to 285.93 m
(with the 300 m
range
selected)
(If the 3
range is selected, the threshold is set to 2.8593
)
Query
:CALC:LIM:RES:UPP?
Response
28593
Note
The value is sent as a whole integer (count). To set 120.53 m
with the 300 m
range, send the following:
:CALC:LIM:RES:UPP 12053
(Voltage Measurement)
Syntax
Command
:CALCulate:LIMit:VOLTage:UPPer
<Upper threshold>
Query
:CALCulate:LIMit:VOLTage:UPPer?
Response
<Upper threshold>
<Upper threshold>
= 0 to 999999
(NR1)
Example
Command
:CALC:LIM:VOLT:UPP 380000
Sets the upper threshold to 3.80000 V
(with the 6 V range selected)
Query
:CALC:LIM:VOLT:UPP?
Response
380000
Note
The value is sent as a whole integer (count).
To set 48.5003 V with the 60 V range, send the following:
:CALC:LIM:VOLT:UPP 485003
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...