4.1 Comparator Function
43
4
C
ha
pte
r 4
A
pp
lie
d M
ea
su
re
m
en
t
This example describes the comparator setting method.
Comparator Setting Example 1 (Upper and Lower Threshold Judgment)
Example:
Set the upper and lower thresholds for resistance and voltage in the
V mode (300
m
range), and indicate whether the measurement value exceeds the upper or
lower thresholds by sounding the beeper.
Resistance
: Upper threshold value 150.00 m
, Lower threshold value 100.00 m
Voltage
: Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
1
Confirm that the Comparator function is OFF.
First make sure the Comparator function is disabled. Settings cannot be changed while
the Comparator function is enabled. Press the
COMP
key, if necessary, to disable the
Comparator function.
2
Select the
V measurement mode.
3
Select the Resistance measurement range (for this example, the 300 m
range).
4
Select the voltage measurement range (for this example, the 60 V range).
COMP not lit
V lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
m
lit
V lit
Increase the resistance measurement range.
Decrease the resistance measurement range.
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...