![Hioki BT3562 Instruction Manual Download Page 185](http://html1.mh-extra.com/html/hioki/bt3562/bt3562_instruction-manual_2137240185.webp)
8.8 Sample Programs
179
8
C
ha
pte
r 8
R
S-2
32
C
/G
P-IB
/L
AN
In
te
rfa
ce
s
(a) This makes it so that during communication the [ Begin Measurement ] and [ Close ] buttons cannot be
pressed.
(b) Matches communication conditions and the computer usage conditions.
The port to be used on the computer: 1
Transmission speed: 9600 bps
Parity: none
Data length: 8 bit
Stop bit: 1bit
(c) Sets CR + LF as the terminator indicating the end of the sending and receiving character string.
(d) Sets the reading operation time to 2 seconds.
(e) Opens the “data.csv” file. However, if a file with this name already exists, the previous “data.csv” will be
deleted and a new file created.
(f) Sends the command to perform one measurement and return that measurement result to the com
-
puter.
HIOKI BT3562A981-09