4.1 Comparator Function
50
11
Press so that the indicated position blinks, and select voltage.
r
.............Resistance
u
............Voltage
12
Press so that the indicated position blinks, and select the comparison method for
the comparator (here, REF/%).
HIGH, LOW
.....Compare by upper and lower thresholds (default setting)
REF, %
.............Compare by reference value and tolerance
13
Switch to the Ref/% threshold setting display, and specify the thresholds.
14
Applies setting and returns to the Measurement display.
The comparator function is enabled.
To cancel the settings:
SHIFT
key
u flashing
REF & % flashing
Or
numeric
keypads
For this example,
Reference value: 4.2 V
Tolerance: 0.5%
COMP lit
HIOKI BT3562A981-09