8.6 Message Reference
158
Execute a Measurement and Read the Measured Values
Syntax
Query
:READ?
Response <Measured resistance
(NR3)
>, <Measured voltage
(NR3)
>
(
V mode)
<Measured resistance
(NR3)
>
(
mode)
<Measured voltage
(NR3)
>
(V mode)
Description
Query
Switches from the Idle State to the Trigger Wait State, then reads the
next measured value. With auto-ranging enabled, the most suitable
range is selected before measurement.
Example
Query
:READ?
Response
289.68E-3, 1.3921E+0
(
V mode)
Measured resistance is 289.68 m
, and voltage is 1.3921 V.
Error
This command causes an execution error if issued during the Continuous Mea
-
surement state (after
:INITIATE:CONTINUOUS ON
).
Note
• The next command does not execute until measurement is finished.
• When the trigger source is external, the
TRG
command does not trigger mea
-
surement.
• Wait 100 ms before applying the trigger with
:READ?
immediately after
changing the measuring conditions during measurement.
See
" Measurement Value Formats" (p.159).
Trigger Source Operation
IMMediate
Triggers and reads measured value.
EXTernal
After triggering by the TRIG terminal (EXT I/O) or
TRIG
key,
reads the measured value.
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...