9.2 Accuracy
187
9
C
ha
pte
r 9
S
pe
cific
atio
ns
Hioki expresses accuracy as error limit values specified in terms of percentages relative to reading and full
scale, and digits.
Resistance Measurement
*1: BT3562A, BT3562, BT3563A, and BT3563 only
*2: Measurement current error within ±10%
*3: Add ± 3 digits for EX.FAST, or ± 2 digits for FAST and MEDIUM sampling rates.
Add ±
30 digits for EX.FAST, or ± 10 digits for FAST, or ± 5 digits for MEDIUM sampling rates (3 m
range).
Voltage Measurement
*4: BT3562A only
*5: BT3563A and BT3563 only
*6: Add ± 3 digits for EX.FAST, or ± 2 digits for FAST and MEDIUM sampling rates.
Warm-up time
At least 30 minutes
Self calibration
Except when using SLOW sampling, self-calibration should be executed
after warm-up.
Ambient temperature after self-calibration should be maintained within ± 2
°
C.
Measurement state
Measurement taken in the same set-up as was in place when zero adjust
-
ment was performed, including identical probe profile and placement.
Probe profile must not be changed during measurement.
Full scale
(Maximum displayable value)
Refers to the maximum displayable value of each measurement range.
The instrument has measurement ranges whose values are equal to the maximum displayable values.
The limit values of full-scale errors are expressed in percent of full scale (% of full scale, % f.s.).
Reading
(Displayed value)
Refers to the displayed value of the measuring instrument.
The limit values of reading errors are expressed in percent of reading (% of reading, % rdg.).
Digits
(Resolution)
Refers to the smallest change in the indication on the digital measuring instrument, i.e., the numeral one in the
rightmost place. The limit values of digit errors are expressed in terms of digits (dgt.).
Guaranteed Accuracy Conditions
Range
3 m
*1
30 m
300 m
3
30
300
3000
Maximum
displayed values
3.1000 m
31.000 m
310.00 m
3.1000
31.000
310.00
3100
.
0
Resolution
0.1
1
10
100
1 m
10 m
100 m
Measured current
*2
100 mA
100 mA
10 mA
1 mA
100
A
10
A
10
A
Measured current
frequency
1 kHz ±
0.2 Hz
Accuracy
*3
± 0.5% of reading ±
5 digits
± 0.5% of reading ±
10 digits (3 m
Range)
Temperature
coefficient
(
±
0.05% of reading ±
0.5 digits)/°C
(±
0.05% of reading ±
1 digit)/°C (3 m
Range)
Range
6 V
60 V
100 V
*4
300 V
*5
Maximum displayed
values
±
6.00000 V
±
60.0000 V
±
100.000 V
±
300.000 V
Resolution
10
V
100
V
1 mV
1 mV
Accuracy
*6
±
0.01% of reading ±
3 digits
Temperature coeffi
-
cient
(±
0.001% of reading ±
0.3 digits)/°C
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...