4.1 Comparator Function
58
In
V mode, both resistance and voltage measurement values are displayed.
Although comparator setting values are not normally displayed when the com
-
parator is enabled, they can be displayed for confirmation by the display switch
-
ing function.
Switching Between Measurement Value and Comparator Setting Displays
Example:
Resistance: Upper threshold value 150.00 m
, Lower threshold value 100.00 m
Voltage: Upper threshold value 15.2000 V, Lower threshold value 15.0000 V
Press this key to switch the display between measurement values
and comparator setting values.
Measurement display switching is available only with the comparator
enabled, and in the
V mode.
Use it to confirm comparator setting values.
Resistance and voltage measurement display
(Shows resistance and voltage measurement values simultaneously)
Resistance measurement and comparator
display (Shows resistance measurement and
resistance comparator setting values)
Voltage measurement and comparator dis
-
play (Shows voltage measurement and volt
-
age comparator setting values)
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...