8.5 Message List
123
8
C
ha
pte
r 8
R
S-2
32
C
/G
P-IB
/L
AN
In
te
rfa
ce
s
Statistical Functions
:CALCulate:STATistics:STATe
1/ 0/ ON/ OFF
Sets statistical calculation function ex
-
ecution
:CALCulate:STATistics:STATe?
ON/ OFF
Queries the statistical calculation
function execution setting
:CALCulate:STATistics:CLEAr
Clears statistical calculation results
:CALCulate:STATistics:RESistance:NUMBer?
<Total data count>,
<Valid data count>
Queries the resistance data count
:CALCulate:STATistics:VOLTage:NUMBer?
<Total data count>,
<Valid data count>
Queries the voltage data count
:CALCulate:STATistics:RESistance:MEAN?
<Mean>
Queries the resistance mean value
:CALCulate:STATistics:VOLTage:MEAN?
<Mean>
Queries the voltage mean value
:CALCulate:STATistics:RESistance:MAXimum?
<Maximum value>,
<Data No. of Maximum
value>
Queries the resistance maximum val
-
ue
:CALCulate:STATistics:VOLTage:MAXimum?
<Maximum val
-
ue>,<Data No. of Maxi
-
mum value>
Queries the voltage maximum value
:CALCulate:STATistics:RESistance:MINimum?
<Minimum value>,
<Data No. of Maximum
value>
Queries the resistance minimum val
-
ue
:CALCulate:STATistics:VOLTage:MINimum?
<Minimum value>,
<Data No. of Maximum
value>
Queries the voltage minimum value
:CALCulate:STATistics:RESistance:LIMit?
<Hi count>,<IN count>,
<Lo count>,
<Measurement fault
count >
Queries comparator results of resis
-
tance measurement
:CALCulate:STATistics:VOLTage:LIMit?
<Hi count>,<IN count>,
<Lo count>,
<Measurement fault
count >
Queries comparator results of voltage
measurement
:CALCulate:STATistics:RESistance:DEViation?
<
n
>, <
n-1
>
Queries standard deviation of resis
-
tance measurement
:CALCulate:STATistics:VOLTage:DEViation?
<
n
>, <
n-1
>
Queries standard deviation of voltage
measurement
:CALCulate:STATistics:RESistance:CP?
<Cp>, <CpK>
Queries process capability indices of
resistance measurement
:CALCulate:STATistics:VOLTage:CP?
<Cp>, <CpK>
Queries process capability indices of
voltage measurement
Memory Function
:MEMory:STATe
1/ 0/ ON/ OFF
Sets the memory function state
:MEMory:STATe?
ON/ OFF
Queries the memory function state
:MEMory:CLEAr
Clears instrument memory
:MEMory:COUNt?
0 to 400
Queries the memory data count
:MEMory:DATA?
[STEP]
Queries the memory data
Self-Calibration
:SYSTem:CALibration
Executes self-calibration
:SYSTem:CALibration:AUTO
1/ 0/ ON/ OFF
Sets automatic self-calibration
Message ([ ] = optional)
Data Contents
( ) = response data
Description
Ref
page
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...