10.2 Cleaning
191
10
C
ha
pte
r 1
0 M
ain
te
na
nc
e a
nd
S
erv
ic
e
To clean the instrument, wipe it gently with a soft cloth moistened with water or
mild detergent. Never use solvents such as benzene, alcohol, acetone, ether,
ketones, thinners or gasoline, as they can deform and discolor the case.
10.2 Cleaning
10.3 Error Display
Display
Description
Err02
Zero-Adjust Range Error
The resistance measured value or voltage measured
value prior to zero adjustment exceed 1000 digits
Err10
Execution Error
The data portion of a remote command is invalid.
Using the :READ? command with the :INITIATE:CON
-
TINUOUS command set to ON will also cause the
Err10 error.
Set the :INITIATE:CONTINUOUS command to OFF
before executing the :READ? command.
Err11
Command Error
The command portion of a remote command is invalid.
Err90
ROM Error
An internal program error occurred. Repair is required.
Err91
RAM Error
An internal RAM error occurred. Repair is required.
Err92
EEPROM (Adjustment Data)
Error
Adjustment data is corrupted. Repair is required.
Err95
A/D Communications Error
The A/D converter is damaged. Repair is required.
- - - - -
This indicates a measurement fault. It appears in cases of a disconnected test lead, poor
probe contact or when the test object’s measured value is far above the measurement
range.
The measurement fault signal is output from the ERR terminal of the EXT I/O connector.
The following causes should be considered:
• A test lead may not be connected to the test object
• Test object resistance may be too very large for the measurement range
Example: Measuring 20
with the 300 m
range
• Any of the SOURCE-H, SOURCE-L, SENSE-H or SENSE-L leads may be discon
-
nected or poorly connected
• The probe may have a high contact resistance
See
" Measurement Fault Detection"(p.39)
• The contact failure circuit protection fuse may have blown due to test lead damage,
excessive wear, or impurities.
HIOKI BT3562A981-09
Summary of Contents for BT3562
Page 2: ...HIOKI BT3562A981 09...
Page 24: ...1 5 Measurement Flowchart 18 HIOKI BT3562A981 09...
Page 84: ...4 13 Reset Function 78 HIOKI BT3562A981 09...
Page 102: ...6 3 Printing 96 HIOKI BT3562A981 09...
Page 186: ...8 8 Sample Programs 180 HIOKI BT3562A981 09...
Page 198: ...10 3 Error Display 192 HIOKI BT3562A981 09...
Page 200: ...194 HIOKI BT3562A981 09...
Page 220: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3562A981 09...
Page 224: ...Index HIOKI BT3562A981 09...
Page 225: ...HIOKI BT3562A981 09...
Page 226: ...HIOKI BT3562A981 09...
Page 227: ...HIOKI BT3562A981 09...
Page 228: ...HIOKI BT3562A981 09...