10.1 Using the 8910 with the 8826 MEMORY HiCORDER
106
Follow the procedure below to send data.
1.
Connect the 8826 to the 8910 with the RS-232C cable.
2.
Position the [
ONLINE/OFFLINE
] switch of the 8910 to [
ONLINE
] side.
3.
On the 8910 setting screen of the 8826, select a block to which to
send data, then click [
F5
] (RECORDER
→
8910 function key).
4.
After the self-check function checks connectivity, the 8826 sends data
in following order:
(1) The registration list (up to 50 items)
(2) Analog CH (12 channels)
(3) Logic CH (24 bits)
The data send status is indicated in the communication message
display section at the lower left of the screen.
5.
Once the data is transmitted, the communication message display
section is cleared and communication terminates.
If data transmission fails, the communication message display section
indicates "
Can not connect.
"
.
Sending data to the 8910
Communication message display section
Block No.
If the error massage
appears
Check the cable for proper connection, and
that the [
ONLINE/OFFLINE
] switch of the
8910 is positioned to [
ONLINE
] side.
Summary of Contents for 8910
Page 1: ...8910 Instruction Manual CAN ADAPTER EN Aug 2018 Revised edition 7 8910A981 07 18 08H ...
Page 2: ......
Page 8: ...Contents vi ...
Page 18: ...Notes on Use 10 ...
Page 44: ...3 5 Operation Map 36 ...
Page 48: ...4 2 Program Setup 40 ...
Page 56: ...5 4 Operation Flowchart 48 ...
Page 78: ...6 6 Editing in Another Window 70 ...
Page 90: ...7 4 Setting the ID Filter 82 ...
Page 99: ...91 9 2 Setting Data Printout ...
Page 130: ...10 2 Using the 8910 with the 8841 8842 MEMORY HiCORDER 122 ...
Page 147: ......
Page 148: ......
Page 149: ......
Page 150: ......