![Hioki 3532-50 Instruction Manual Download Page 75](http://html1.mh-extra.com/html/hioki/3532-50/3532-50_instruction-manual_2137204075.webp)
61
_____________________________________________________________________________________________
4.8 Short Circuit Compensation
______________________________________________________________________________________________
4.8.2 Setting the Compensation Method
Metallic plate
Metallic wire
Not possible
Possible
NOTE
NOTE
1.
A shorting bar is used. This shorting bar is for short circuiting together the
ends of the test leads. Use an object whose impedance is as low as possible.
If you use a metallic wire or the like as a shorting bar, try to ensure that it is
as thick and short as possible.
2.
Arrange the test leads as closely as possible to their configuration in which
measurement will be performed, and short circuit together the HIGH and
LOW leads. In order to keep external influences as low as possible, be sure
to thrust the shorting bar in all the way.
If you intend to short circuit between the clamps at the ends of the test
leads,
clip both clamps onto a short piece of metallic wire
as shown.
When using 4-TERMINAL PROBE 9140, please pinch the short wire
with both clips. A short circuit state can not be created by pinching clip
each other.