![Hioki 3532-50 Instruction Manual Download Page 41](http://html1.mh-extra.com/html/hioki/3532-50/3532-50_instruction-manual_2137204041.webp)
27
_____________________________________________________________________________________________
3.3 Basic Measurement
______________________________________________________________________________________________
Short Circuit Compensation Screen
Confirmation Screen
NOTE
Data Determination Screen
4.
Press the
key so as for example to select
ALL compensation, with which short circuit
compensation values for all the test
frequencies will be determined and adjusted.
5.
A Confirmation screen will be displayed for
confirmation of ALL short circuit
compensation. If it is OK to proceed, press
the
key.
Check again - are the test leads properly short
circuited together by the shorting bar?
6.
When you press the
key, the ALL short
circuit compensation process starts.
After about 2 minutes, the determination and
adjustment of the short circuit compensation
values is complete.
If an error message appears, refer to Section
4.8.5.