139
_____________________________________________________________________________________________
5.6 Testing Using EXT I/O
______________________________________________________________________________________________
5.6.2 Test on the Continuous Test Screen
TRIG
(Test start signal)
INDEX
Analog testing signal)
EOM
(Test end signal)
M-IN
S-IN
AND
(Comparator result
output
Setting of the
panel number 6
Setting of the
panel number 8
Setting of the
panel number 10
Decision
result
Previous decision result
NOTE
On the Continuous test screen, when a trigger signal is input via the EXT
I/O connector, or when on the screen the
soft key is pressed, then
the decision result is output on the comparator result output signal line of
the EXT I/O connector after testing is completed for all the panel numbers
that must be tested on the screen according to the setting.
When performing continuous testing with panel numbers 6, 8, and 10 for
example.
AND
__________
If all of the decision results are IN, then LO is output.
M-IN
___________
If all of the decision results of the first parameters are IN, then LO is output.
S-IN
_________
If all of the decision results of the third parameters are IN, then LO is
output.
M-HI
___________
, M-LO
___________
, S-HI
__________
, S-LO
___________
: No output
On the continuous testing screen, the panel load signals (LD1 to LD5)
cannot be used.
For details on continuous testing, see Section 4.22, "Continuous Test
Function."